Drop volume | 2.5 nL |
Volume transfer range |
2.5 nL – 5 µL (dependent on fluid type, assay, and source plate used) |
Reagents supported |
DMSO,
PCR and qPCR reagents, synthetic biology reagents, NGS reagents,
primers and probes, cell culture media, proteins, nucleic acids, up to
50% glycerol, serum, plasma, antibodies, enzymes, and protein
crystallography reagents |
Transfer accuracy |
<10% deviation from target volume |
Transfer precision |
<8% CV (except protein crystallography fluid class) |
Source labware compatibility |
All Echo Qualified 384- and 1536-well Microplates |
Destination labware compatibility |
All
Echo Qualified Microplates and most ANSI-compliant/SBS-standard
microplates in 96-, 384-, 1536-, and 3456-well formats 8 to 16 mm well
height |
Barcode reader options |
Short-side and/or long-side |
Dimensions |
53.9 cm width x 68.3 cm depth x 92.5 cm height (21.2 x 26.9 x 36.4 inches) |
Weight |
130 kg (286 lbs) |
Working envelope |
Additional 25.4 cm (10 inches) on top, 2.5 cm (1 inch) on sides, 7.6 cm (3 inches) at front, 30.5 cm (12 inches) at back |
Electrical |
AC 120V, 50/60 Hz, 10 A or AC 230V, 50/60Hz, 5 A |
Operating temperature |
21°C ± 5°C (70°F ± 9°F) |
Vacuum |
House supply minimum 200 Torr (266 mbar, 22 inches Hg Vac)
OR
Standalone vacuum with surge tank, minimum pump speed 2.0 m3/hr, minimum pressure 100 Torr (133 mbar, 26 inches Hg Vac) |
Fluid class calibration | No calibration needed |