technology

right Pharmacy See all product

PTB 111EP Manual Tablet Hardness Testing Instrument with Integrated Report Printer

☆☆☆☆☆ ( 0 rates ) 285
Price : Call

Code: #22327

Supplier: Pharma Test Apparatebau AG

PTB 111EP manual tablet hardness testing instrument with integrated printer available in different hardness ranges

rightMORE INFORMATION

  • The manual tablet hardness testing instrument PTB111EP / 111EP-500 / 111EP-800 features an integrated report printer. It is a dual force mode hardness test apparatus as it can be used for either linear force or linear speed increase while tablet hardness is tested. It offers a multiple point validation procedure for the built-in digital load cell. The instrument is made in strict compliance with the EP <2.9.8> and USP <1217> Pharmacopoeia.
     
    Features
    Fully USP <1217> and EP <2.9.8> compliant
    Dual force mode instrument with linear speed increase and linear force increase modes
    Steppless adjustment of the linear force or speed increase rate
    Integrated thermal printer
    Multiple point validation procedure built-in
    Programmable print-out of force increase curve
    Set to test tension strength of oblongs and caplets available

  • Parameter

    Specification

    Hardness Testing Range

    PTB111EP: 5.0 – 300.0 N ± 2 N

    PTB111EP-500: 10.0 – 500.0 N ± 5 N

    PTB111EP-800: 10.0 – 800.0 N ± 5 N

    Hardness Accuracy

    Better than ±1 N

    Hardness Resolution

    0.1 N

    Measuring Units

    Newton (N), Kilopond (kp) and Strong Cobb (Sc)

    Force Mode

    Selectable: linear force increase or linear speed increase

    Force Rate

    (linear force increase)

    PTB111EP, PTB111EP-500: 5 – 200 N/sec

    PTB111EP-800: 10 – 200 N/sec.

    Force Rate

    (constant speed)

    5 – 200 mm/Min

    Breakpoint Detection Range

    Min. 2mm, Max. 10 mm

    Sample Touch Force

    Min. 48 digits (approx. 3N), Max. 999 digits

    Breakpoint Detection Force

    Min. 48 digits (approx. 3N), Max. 999 digits

    Internal Printer

    Thermo-Printer Miniplus-P3B, parallel, 58 mm paper width, 30 mm paper roll diameter

    Interface

    RS-232 serial, data transfer

    USB PCL 3-5 type printers only by use of a LPT to USB adapter (part no. 002-0461)

    Instrument Housing

    Stainless steel (304) to meet GLP requirements

    Power

    115 Volt AC or 230 Volt AC, 50/60 Hz

    Current Consumption

    0,15 A at 230 V; 0,30 A at 115 V; 35 W

    Main Fuse

    2 x T1,0 A

    Installation Requirements

    Ambient Temperature 15-35 °C

    Relative Humidity 15-80 %rH

    Desk with at least 50 kg working load. All around the instrument at least 10 cm free distance to walls or other equipment. Free access to the mains power plug and switch.

    Instrument Dimensions

    Approx. 260 x 500 x 260 mm (Length x Width x Height)

    Packaging Dimensions

    Approx. 310 x 660 x 420 mm (Length x Width x Height)

    Net / Gross Weight

    Approx. 9,5 kg / 12,5 kg (without optional accessories)

    Certification

    All components certified to USP / EP requirements

    CE / EMC Certification

    All CE / EMC Certification provided

    Validation

    All IQ & OQ documents included

  • No external printer required, this reduces the necessary bench space
    Documentation of all results using an integrated report printer
    Report printer uses less paper compared to a standard A4 format printer
    Select either linear force or linear speed increase (dual mode selection)
    Entry of time and date
    Enter a 12 digit batch number
    Automatic re-start facility to speed up the testing sequence
    Validation and calibration program for the measurement station
    Force curve print out
    Dual point adjustment of the load cell for the hardness test station
    Multiple point validation (calibration)
    Programmable print-out of force increase curve
    Data transfer via RS-232 interface
    Hardness testing in compliance with the EP <2.9.8> and USP <1217> Pharmacopoeia
    Test program for soft gelatine capsule testing by setting up a testing distance

 
Scroll