technology

right Bio-Technology See all product

PMS LiQuilaz II S03 Particle Counting Spectrometer for Liquids (0.3 – 3.0 µm)

☆☆☆☆☆ ( 0 rates ) 179
Price : Call

Code: #21358

Supplier: Particle Measuring System

PMS LiQuilaz II S03 Particle Counting Spectrometer for Liquids (0.3 – 3.0 µm)

rightMORE INFORMATION

  • Applications
    - Effervescent chemical measurements using compression sampling
    - Chemical quality assurance
    - Process bath monitoring
    - DI water monitoring
    - Chemical delivery systems
    - Photoresist monitoring

  • - Size range 0.2 – 2.0 µm
    - Channels User adjustable from 1 to 15
    - Flow rate: 80 ± 10% ml/min
    (20 ± 10% mL/min available for all S series models for sampler applications)
    - Sample volume: Volumetric 80 ml/min (100%)
    - Maximum concentration: 10,000 particles/ml
    - Sample temperature: 32 – 302 °F (0 – 150 °C)
    - Zero count: < 15 counts/liter
    - Pressure range: 0 - 100 psi
    - Laser source: Laser diode
    - Wetted surface materials: Sapphire, Teflon®, Kalrez® 4079 and Kel-F®
    - Dimensions (l, w, h): 32.4 x 10.8 x 11 .4 cm
    - Weight: 6.1 lb (2.8 kg)
    - Electrical rating: 100 - 240 V, 50/60 Hz, 0.75 A
    - Communication modes: 5-wire, RS-485, DE9 connectors, PMS serial communications protocol, 10/100BaseT Ethernet, PMS
    Ethernet Protocol, Modbus TCP
    - Calibration: Materials used are traceable to U.S. National Institute of Standards and Technology (NIST) and/or
    Japanese Institute of Standards (JIS)
    - Environment: Temperature: 50 – 95 °F (10 – 35 ºC); Humidity: non-condensing

 
Scroll