technology

right Electronic and Information Technology See all product

Oil in Water Analyser (MS1200)

☆☆☆☆☆ ( 0 rates ) 24
Price : Call

Code: #28848

Supplier: WTE Co., Ltd

Oil in Water Analyser (MS1200)

rightMORE INFORMATION

  • The MS1200 on-line Oil in Water and Pollution Analyser is designed to protect drinking water treatment plants from pollution events at the raw water intake. These events can result in expensive filter replacement and clean-up operations and may also affect the quality of drinking water produced.
     
    The MS1200 utilises a contactless measurement technique based on Henry’s Law, sensing headspace gases or volatiles in the sampling tank provided, and provides a measurement system with very low maintenance requirements.
     
    Thanks to the clever design and technology the sensor never touches the water. This results in reduced drift, virtually no fouling, no sensor cleaning and very low maintenance.
     
    The MS1200 is accurate to low ppb concentrations and its wide dynamic range allows it to be used in a wide variety of environments.
     
    The concentration of volatile organic compounds (VOCs) and water vapor in the headspace increases until equilibrium is reached with the concentration in the water. A sample of gas is then taken from the headspace and the response analysed to provide a concentration in ppb of contaminants in the water.
     
    APPLICATIONS
    • Monitoring of water abstraction points
    • Monitoring of drain and storm water systems
    • Detection of fuel pollution in surface water
    • Detection of VOC breakthrough in carbon beds
    • Reverse Osmosis Membrane protection
    • Protection of desalination plants 

  • Product Name

    MS1200

    Measuring Items

    Oil in water : VOC, Fuel oil, BTEX, PAH, etc.

    Manufacturer

    Multisensor Systems

    Detection Range

    1 ~ 3000 ppb

    Display Range

    0 ~ 1000 ppb

    Repeatability

    -2 ~ +2%

    Accuracy

    -10 ~ +10%

    Measurement Frequency

    Continuous (15, 20, 30, 60 minutes modes available on
    request)

    Ambient Temperature

    0 ~ 40°C

    Water Temperature

    1 ~ 40°C

    Analogue Output

    4 ~ 20mA (Scalable to range required, max load 900Ω)

    Analogue Output Isolation

    400V DC

    Relay Voltage

    50V (3x: Alarm 1, Alarm 2 and Fault Relays with NO and NC contacts)

    Relay Current

    5A

    Service Interface

    USB-A to PC (Using Multisensor Software provided for non touchscreen version)

    Data Storage

    μSD Card (6 months)

    Instrument Case

    IP65 (Coated Mild Steel)

    Instrument Weight

    16kg

    Sampling Chamber Material

    Stainless Steel (Optional: PVC)

    Sampling System Weight

    12kg (Without water)

    Sampling System Dimensions

    570 x 490mm

    Sampling System Capacity

    3 litres

    Flow Limit Switch

    Contacts closed if flow below set point (Option available on request)

    Water Flow Rate

    2 L/min

    System Dimensions

    1170 x 490 x 300mm (Mounted on 2 separate PVC backboards)

    Supply Voltage

    90 ~ 240V AC (50Hz or 60Hz)

    Power Consumption

    Standard Version: 15W (Typical 10W durig operation)
    Touch Screen Version: 45W (Typical 20W during operation)

  • • Accurate to low ppb concentrations
    • Contactless measurement
    • Not affected by turbidity
    • Detects fuel oils, PAH, VOCs and BTEX
    • Continuous measurement
    • Continuous monitoring
    • Low operating and maintenance costs
    • Monitors for pollution events
    • Low maintenance, no sensor cleaning
    • Insensitive to turbidity
    • High sensitivity, ideal for boreholes

 
Scroll