Type 2 water using reverse osmosis and deionisation technology
- 1-15 MΩ-cm
- Type 2 Water Quality general laboratory grade water
- >50% recovery of RO water
- Novel recirculation process extends life of RO membrane and DI cartridges
- TOC <50ppb
- Operates at 0.1-6 bar feed water pressure
- Up to 98% rejection of inorganics
- Compact design
- Wall mounting brackets included
- Bench, reservoir-top or wall mounting
- Excellent monitoring of conductivity
- SD Card interface
- Choice of 30/60/100 litre capacity reservoir
PRODUCT WATER QUALITY
MODEL
|
DI MAKE UP RATE (@15°C)
|
INORGANICS REJECTION RATE**
|
DI QUALITY (@25°C)
|
DISPENSE RATE (@15°C FROM TANK)*
|
TYPICAL TOC REDUCTION***
|
GENO10 |
10L/hr |
98% |
1 - 15MΩ-cm |
3 - 7L/min |
96% |
GENO20 |
20L/hr |
98% |
1 - 15MΩ-cm |
4 - 7L/min |
96% |
GENO50 |
50L/hr |
98% |
1 - 15MΩ-cm |
5 - 7L/min |
96% |
* Depending on tank and pump setup
** Depending on feedwater impurities
*** Using tank UV option
SYSTEM REQUIREMENTS
FEED WATER
|
|
|
Pressure |
Bar |
0.1 - 6 |
Conductivity |
μS/cm |
<1400 |
Temperature |
°C |
4 - 30 |
Hardness as CaCO3 |
ppm |
<400 |
Free Chlorine |
ppm |
<2 |
Colloid Index |
SDI |
<3 |
pH Value |
|
4 - 10 |
Iron (Total Fe) |
ppm |
<0.1 |
UNIT SPECIFICATIONS
ALL MODELS
|
|
Dimensions (HxWxD) |
500 x 490 x 290mm |
Installed Weight |
23kg |
Power Supply |
100 - 240V ~; 50/60Hz; 3.0A (max) |
Power Supply Output |
24VDC, 5.OA, 120W (max) |
Operating Temperature |
4°C - 35°C (max) |
Storage Temperature |
4°C - 35°C (max) |
UTILITIES & SERVICES
REQUIREMENTS
|
|
Water Supply |
Isolated cold-water feed |
Drain Type |
Sink/Tundish/Standard up-stand with suitable trap |
Drain Flow |
>3L/min |
CONSUMABLES & ACCESSORIES
PART NUMBER
|
DESCRIPTION
|
TYPICAL ANNUAL REQUIREMENT
|
TC001 |
Pre-treatment Module |
2 |
TC002 |
Endure™ Purification Pack 1 |
2 |
TC005 |
Tank Vent Filter |
1 |
TANK30 |
30 Litre Tank |
- |
TANK60 |
60 Litre Tank |
- |
TANK100 |
100 Litre Tank |
- |
TANK100P-1 |
100 Litre Tank with Pump - 230V |
- |
TANK100P-2 |
100 Litre Tank with Pump - 110V |
- |
AV004 |
Valve Kit for Washing Machine Feed (Single POU) |
- |
AV006 |
Wall Mounting Kit |
- |
AV007 |
External Pre-filter Housing Kit (Excludes Filter) |
- |
AV016 |
Valve Kit for Washing Machine Feed (Double POU) |
- |
TYPE 2 WATER
APPLICATIONS
|
Buffer & Media Preparations |
Sample Dilution & Reagent Preparation |
Spectrophotometry |
Protein Electrophoresis |
Cytology & Histology |
Glassware Washing & Rinsing |
RESERVOIRS
|
WIDTH
|
HEIGHT
|
DEPTH
|
TANK30
|
410mm
|
500mm
|
240mm
|
TANK60
|
485mm
|
500mm
|
330mm
|
TANK100
|
485mm
|
800mm
|
340mm
|
TANK100PUMP*
|
485mm
|
800mm
|
340mm
|
*100 litre tank with built-in pump to provide the correct
pressure for feeding washing machines, glasswashers and remote points of
use.
- Including dispense tap on front of the reservoir in the middle or at the bottom
- 3/4” BSP glass-washer outlet
- Pressure sensor level control, with a 0-100% contents display
- 0.2μm microbiological air vent filter or an optional CO2 filter
- 30 and 60 litre reservoirs are wall mountable
SERVICES REQUIRED
3/4” BSP isolated cold water feed (washing machine tap/valve)
240 Volt, 13 Amp sockets
Drain/up-stand with trap or sink/tundish
OPTIONAL LEAK DETECTION
If your water system is in a location where a leak could cause damage
or health and safety issues, we recommend that you protect against this
possibility by means of a leak detector, or for larger systems a
containment bund. Containment bunds, price available on application. The
following are available from Avidity Science:
PART NO.
|
DESCRIPTION
|
2084
|
Water Guard for small laboratory equipment up to 100L/hr
|
00029
|
Water Guard for larger equipment with a feed connection up to ¾” |