technology

Gas Volume Analyzer GVA-710

☆☆☆☆☆ ( 0 rates ) 109
Price : Call

Code: #29726

Supplier: Kyoto Electronics Manufacturing Co., Ltd.

This instrument automatically measures gas volume, internal gas pressure, air content and oxygen concentration for quality control in the bottling process of carbonated drinks, etc.

rightMORE INFORMATION

  • Soft drinks and brewery
    Capable of automatically measuring gas volume / internal gas pressure, air content and oxygen concentration for quality control in the bottling process of soft drinks, etc.
     

    Measurement items

    Applications

    Gas volume / internal gas pressure (volume of carbon dioxide)

    Carbonated drink
    (PET bottles, metal cans and glass bottles)

    Air content (Residual gas volume) with optional oxygen concentration

    Carbonated drink
    (PET bottles, metal cans and glass bottles)

    Inside gas pressure

    Carbonated / non-carbonated drink
    (PET bottles, metal cans and glass bottles)

    Supported standards
     

    Control items
    Measuring items

    Gas volume
    (internal gas pressure)

    Air content

    Gas volume
    + air content
    (simultaneous measurement)

    Oxygen
    concentration
    Option*1

    Inside
    gas
    pressure

    Soft drinks
    Carbonated drink

    A

    A

    A

    S

    A

    Soft drinks
    Carbonated fruit juice

    S

    S

    S

    S

    S

    Soft drinks
    Fruit juice

    -

    -

    -

    -

    S

    Soft drinks
    Coffee

    -

    -

    -

    -

    A

    Brewery
    Beer

    A

    A

    A

    S

    A

    Brewery
    Cocktail with tonic water

    A

    A

    A

    S

    A

    Brewery
    Cocktail with fruit juice

    S

    S

    S

    S

    S

    Brewery
    Local beer

    A

    A

    A

    S

    A

     

  • Type

    Gas Volume and Air Content Analyzer

    Model name

    GVA-710

    Measurement items

    1)Carbonated beverages:
    Gas volume*1, internal gas pressure, Residual gas volume and its oxygen concentration*2, and air content
    2)Non-carbonated beverages:
    Inside gas pressure

    Measurement range

    1)Pressure: 0.010 to 0.500 MPa
    2)Temperature: 0 to 50

    3)Residual gas volume: 0 to 30 mL
    4)Oxygen concentration: 0 to 21 vol%

    Measurement accuracy

    1)Pressure: ±0.001 MPa
    2)Temperature: ±0.05

    3)Residual gas volume:
    ± 2 %(in case the volume is 5 mL to 30 mL)
    ± 0.1 mL(in case the volume is 0mL to 5mL)
    * When measuring standard gas
    4)Oxygen concentration: ± 0.2 vol%

    Measurement time

    Gas volume/internal gas pressure: Approx. 90 seconds for one sample
    Residual gas volume: Approx. 5 to 12 minutes for one sample
    *It depends on sample and conditions

    Sample container

    Various steel cans, aluminum cans, PET bottles, and glass bottles
    Capacity: Approx. 160 mL to 2.0 L
    Size: height 85 mm to 315 mm, diameter 53 mm to 110 mm

    Stored data

    500 latest measurement results
    20 methods

    External output

    1)RS-232C: 1ch
    for connection to a printer or a computer
    2)USB port: 1ch
    for data transfer and barcode reader

    Ambient condition

    1)Temperature : 5 to 35
    2)Humidity : 85 %RH or below

    Power supply

    AC 100 V to 240 V ± 10 % 50/60 Hz

    Power consumption

    40 W

    Dimensions

    495(W) × 555(D) × 570(H) mm

    Weight

    Approx. 38 kg

    Coating

    Alkali resistant coating

    Supplied air pressure

    0.6 to 0.7 MPa

  • Fully automatic measurement
    - Both gas volume and air content in a sample container can be measured with a single measurement.
    - No personal differences in measurement results by automatic measurement.
    - Less workload compared to manual measurement.
    - Much safer compared to the manual method by less contact to alkaline solutions.
    Reduced measurement and preparation time
    - Reducing time by 30 to 70% compared to the conventional model when measuring air content.
    Support for a variety of samples
    - Enable to measure cans, bottles, and PET bottles from 160 mL to 2.0 L.
    - Enable to measure sample container with tamper-resistant cap (double-layered lid).
    - Enable to measure 500mL PET bottle lightweight containers.
    Output results to external printer or PC using dedicated software
    - Measurement results can be output to not only an external printer but also a connected PC using dedicated software.
    Method function
    - Up to 20 methods depending on sample property and measurement items can be registered in the instrument.
    - The measurement method can be edited on a PC using dedicated software.
    Support for barcode reader
    - Barcode information such as sample ID affixed on the container can be scanned and transferred using a commercial barcode reader
    Visibility inside the absorbent cylinder
    - The dissolution status of carbon dioxide gas and air content in the absorbent cylinder can be visually checked by operators.

 
Scroll