technology

Steam distillation units (S1-S4)

☆☆☆☆☆ ( 0 rates ) 157
Price : Call

Code: #21051

Supplier: behr Labor-Technik GmbH

Areas of use

Food analysisBeverage industryFeed analysisAgricultural industryEnvironmental analysisCosmetics industry

Parameter to be analyzed

AlcoholAmmoniumVolatile acidsFormaldehydePhenol indexProteinNitrogen

Sample material Waste waterAshBaked goodsConstruction additivesBeerSoilsDistillateDieselFertilizing agentFertilizerEssencesExtractsFish and fish productsMeat and meat productsFlue ashFeed and raw materialsGrain and mill productsSpicesHair dyeSludgeCompostPlasticsMaltMilk and milk productsNutritionPerfumeSludgesSparking wineChipboardSpiritsTextilesWaterWineToys

rightMORE INFORMATION

  • Steam distillation unit with automatic addition

    behr steam distillation units are not just high performance units and reliable partners in your daily laboratory routine. An essential factor in the development and design of the devices was also the safety of the user.

    behr distillation units of the S series are the ideal supplement to the behr InKjel digestion systems. Depending on requirements, the user can choose between five steam distillation units. These are identical in the basic construction but differ in terms of operating comfort and degree of automation. The top model behr S 5 is prepared for working with an external titration device.

    • Exemplary safety and reliability
    • Robust and insensitive casing made of polyurethane
    • Distillation time approx. 3 min. per sample
    • Detection limit 0.2 mg N
    • Recovery rate 99.5%
    • Reproducibility ± 1%
    • Adjustable steam generating capacity (10% - 100%)
    • Very easy, menu-guided operation with a single operation element (the behr one-button operation)
    • Programmable reaction time
    • Programmable distillation time
    • USB interface
    • Fill-level monitoring for canister set
    • Practical quick clamping device which the user can operate even "with his eyes closed"


    • Performance capacity and speed
    • Practical and user friendly
    • Exemplary safety and reliability
    • Robust and insensitive casing made of polyurethane
    • Practical quick clamping device which the user can operate even "with his eyes closed"
    • Extremely simple, menu-controlled operation of the programmable units via
    • one single operating element (behr one-button operation).
    • Menu navigation in country language
    • Models S 2 to S 5 have a separate rinsing program







  • S 1 S 2 S 3 S 4
    Item number B00218025 B00233702 B00233703 B00218032
    Packaging quantity 1
    Width (in cm) 42.00
    Depth (in cm) 34.50
    Height (in cm) 70.00
    Weight (in kg) 32.00 34.00 35.00 35.00
    Controller Nein
    Power consumption (in W) 1700
    Current consumption (in A) 9
    Programs 1 1 10 99
    Cooling water consumption (in l/min) 5
    max. distillation time (in min/sample) 2-4 approx. 2-4 min. per sample approx. 2-4 min. per sample approx. 2-4 min. per sample
    Supply container any size; recommended: behrotest® canister set KAS any size; recommended: behrotest® can set KAS any size; recommended: behrotest® can set KAS any size; recommended: behrotest® can set KAS

  • Safe and high-performance steam distillation units

    • Door contact safety switch that is automatically triggered in the event of overload or short circuit
    • Mechanical overpressure safety valve against excessive pressure in the steam generator
    • Vessel monitoring (no distillation possible if no vessel has been inserted)
    • Door contact switch that automatically deactivates the distillation unit if the door is open
    • Resettable overtemperature thermostat (in the case of water shortage in the steam generator)
    • Cooling circuit monitoring by means of pressure switch
    • Temperature-regulated steam heating phase and pressure control via solenoid valve

    behr steam distillation units are not just high performance units and reliable partners in your daily laboratory routine. An essential factor in the development and design of the devices was also the safety of the user

 
Scroll