Control system:
- The K500 model does not have a touch control panel. Control is via an analogue dial.
- Models equipped with a display with touch control system: S100 PCR, K700, K1000, K1300, K1600.
Additional UV lamps:
The
S100 PCR chamber has 2 additional UV lamps in the compression chamber,
increasing the sterility of the air supplied to the working space.
Product description:
- vertical laminar air flow in the range of 0.25 to 0.50 m / s,
- air
flow speed control through a sensor or volt voltage system with the
possibility of +/- 10% correction by the user, low-noise fans,
- HEPA H14 filter and EU3 polyamide pre-filter (G3) according to PN-EN 779 standards,
- automatic compensation of filter wear ensuring constant and safe flow rate in the chambers,
- hour
meter of device operation time and UV fluorescent lamp, permanently
mounted UV germicidal lamp, placed in the upper front part of the
working area, protected against accidental switching on during
operation,
- energy-saving LED fluorescent lamp illuminating the working space,
- stainless steel worktop, single electrical socket on the chamber housing
- optional valves for vacuum, gas and air, and a dedicated base on lockable wheels.
Control
over the correct operating parameters is facilitated by a system that
monitors errors and informs you about the need for a service test or
approaching filter replacement time.