Standards
- The oven performs well inside the apparatus requirements in IEC 60811-2-1 and other equivalent standards.
Temperature range
- +40 °C to +250 °C
Useful volume
- 60 litres
Important specifications
- The sample holder moves up and down by a servo motor
- Special design with controlled air exchange rate and low air speed (<0,001 m/s), which is dependent on the air exchange rate only
- Temperature variation in time ± 0,25 °C
- Temperature variation in space ± 0,5 %
- Temperature controller with 0,1 °C setpoint
- Temperature indicator with sensor in the test space
- Solid state relay for safe control
- Adjustable over and under temperature limits with alarm
- Fixed over temperature fuse at 340 °C
- Over temperature safety relay
- Flowmeter with needle valve, for setting the air exchange rate
- The temperature, timers, alarms and movement are controlled by a micro PLC with colour touch screen
- Laser pointer for measuring the elongation
- Calliper for measuring the tension set
- Temperature controlled cooling fan for the electronics cabinet
- Run-time meter
Included accessories
- Clamps and weights for 8 samples are included as well as the computer (Windows)
- Stylus pen for the instrument touch screen
- Accredited calibration including certificate
- Manual in English
- Spreadsheet templates for testing and result report for
+ 1 material, 3 samples
+ 1 material, 3 samples – round material
+ 2 materials, 2 samples per material
+ 2 materials, 2 samples per material – round material