TOA-DKK CM-31P Handheld Electrical Conductivity Meter (0.1mS/m-10S/m)

rightMORE INFORMATION

  • Measuring Method: AC two-electrode method
    Display: Custom LCD
    Meas. Range:
    [If standard cell is used] Electrical conductivity: 0.1mS/m-10S/m
    Electrical resistivity: 0.1Ω・m-10kΩ・m
    Salinity (NaCl equivalent from electrical conductivity): 0-4.00%
    Temperature: 0-80.0℃
    [If cell for pure water is used] Electrical conductivity: 5μS/m-20mS/m
    Electrical resistivity: 50Ω・m-200kΩ・m
    Temperature: 0-80.0℃
    Display Range:
    +Electrical conductivity:
    0-200.0μS/m
    0-2.000mS/m
    0-20.00mS/m
    0-200.0mS/m
    0-2.000S/m
    0-20.00S/m
    0-200.0S/m
    +Electrical resistivity :
    0.005-2.000Ω・m
    0-20.00Ω・m
    0-200.0Ω・m
    0-2.000kΩ・m
    0-20.00kΩ・m
    0-200.0kΩ・m
    0-2.000MΩ・m
    0-20.00MΩ・m
    +Salinity(NaCl): 0-4.04%
    Temperature: -5-110.0℃
    *In regards to the range, the electrical conductivity/ resistivity differs according to the cell that is used.
    Electrical Conductivity/Resistivity Range Switching: Auto/manual
    Electrical Conductivity/ Resistivity Unit Switching: Can switch between SI Units(S/m, Ω ・ m) and the previous units (S/cm, Ω・cm).
    Repeatability (Main unit):
    +Electrical conductivity: ±0.5%FS
    +Electrical resistivity: ±0.5%FS
    +Salinity ±0.5%FS
    +Temperature: ±0.2℃
    Temperature compensation:
    +Switch setting between Auto/Manual/None Temperature compensation method: Linear/pure water dual temperature compensation
    +Reference temperature: 25℃
    +Temperature coefficient: 0-9.99%(optional setting)
    Calibration: Cell constant calibration
    Temperature Calibration: One-point calibration
    Data Memory: 1000 data points
    Auto Hold Function: Provided (Stability threshold : Fixed)
    Clock Function: Provided (To be shown while conducting a measurement)
    Interval Memory Function: Provided (Interval: The interval can be specified between 1 sec.-99 min. 59 sec. or 2 sec.-99hr. 59 min.)
    Printing Function: Can connect the external printer EPS-P30 (option)
    RS-232C Interface (non-isolated):
    +Connectable Devices: PC or external printer EPS-P30 (option)
    +Communication Specifications: Communication system: Start-stop synchronous method; Baud rate: 19,200 bps;
    Character length: 8 bits; Parity: None; Stop bit: 1 bit
    Analog Output (non-isolated):
    +Number of Outputs/Items: ch1: Number of outputs: 3 Measured value, temperature, and range (only for electrical conductivity/resistivity)
    +Output Specification: Electrical conductivity/ electrical resistivity/ salinity: 0-1V FS (each range); Range: 100mV/range; Temperature: 0-1 V (0-100℃)
    Waterproof Construction: IP67 (Enabled if the sensor is connected and if the external I/O portions are masked) (Can be immersed in water for 1m and 30 min.)
    Performance Compensation Temperature: 0-45℃
    Power Source: AA alkaline battery/nickel-hydrogen battery (2 pieces) or special AC adapter (6VA option)
    Power Consumption(If 3 volt battery is used)*2: Approx. 0.009W
    Battery Life: Approx. 600hours
    Outside Dimensions: Approx. 68 (w) x 35 (h) x 173 (d) mm
    Mass (Including Batteries): Approx. 280g
    Inluding:
    + Main unit
    + Conductivity Sensor: CT-27112B
    + Hand strap: 0TZ00006
    + AA Battery (2pcs)
    + Manual

 
Scroll