DESCRIPTION
Fully automatic, random access, continuous loading, benchtop analyzer for clinical chemistry and immunoturbidimetric assays.
ASSAY TYPES
End Point, Kinetic, Differential, Bichromatic, Fixed Time, Non-Linear. All methods are user programmable (up to 999 tests).
ON LINE REAGENTS
4 racks are provided to hold up to 33 reagents that are divided as follows:
- 2 racks cooled by Peltier cells holding up to 20 reagents
- 2 racks for 13 reagent positions and 14 standards and controls at room temperature
- 2 additional positions for diluents
SAMPLES LOADING
5 racks for continuous loading of samples split in:
- 64 positions for routine samples
- 14 positions for STAT samples
PRIMARY TUBE SIZES
- From 40 mm to 100 mm (height)
- From 10 mm to 16 mm (diameter)
STAT SAMPLE PROCESSING
Allowed at any time.
THROUGHPUT
Up to 333 tests/hour with ISE (200 without ISE).
INCUBATION TEMPERATURE
37 °C.
READING SYSTEM
Direct photometry.
MINIMUM REACTION VOLUME
300 μl.
OPTIC SYSTEM
- Photometer: multi-wavelength, dual channel optic
- Wavelengths: up to 9 narrow band interferential filters 340, 380, 405, 492, 510, 546, 578, 620 (690 opt.)
- Light Source: halogen lamp 6V/10W
- Optical Path: 10 mm
- Linear Range: 0.001 ÷ 2.500 Abs
- Resolution: 0.0001 Abs
SAMPLING ARM
One mechanical arm performs all sampling operations with:
- Capacitive liquid level sensing
- Reagent pre-warming at 37 °C
- Automatic probe washing
DILUTER
Dilutions are made without using traditional syringes with the following specifications:
- Sample volume: 2.0 μl ÷ 99 μl (1 μl incr.)
- Reagent volume: 3.0 μl ÷ 500 μl (1 μl incr.)
REACTION PLATE
Holds 60 semi-disposable cuvettes and perfoms the following operations:
- Automatic cuvettes washing
- Continuous monitoring of cuvettes quality
- Incubation temperature: control at 37 °C ± 0.3
OPTIONS
- Direct potentiometry ISE module for NA+, K+, Cl-
- Positive barcode reader (samples)