SAMPLES AND REAGENTS
Vegasys holds 40 samples (four segments with 10 sample cups each) plus 1 segment for 6 Standards and 2 Controls
Removable reagent tray with 18 reagent containers (25 ml each)
OPERATION
Vegasys runs all clinical chemistry and
immunoturbidimetric tests:End Points, Fixed Time, Kinetics,
Bichromatics and Differential methods with or without Reagent Blank
correction, against Factor or Standard. Samples can be loaded by single
test or Profile. It is programmed for 7 different calculation algorithms
for non-linear analysis, automatic flagging for system and results
anomalies, abnormal values, enzymes substrate depletion, Rbl out of
range.
OPEN SYSTEM
Wide choice of methods and reagents.
The User can programme more than 300 clinical chemistry and
immunoturbidimetric applications.
THROUGHPUT
Up to 200 colorimetric tests/hour.
PHOTOMETER
- Multi wavelength photometer with narrow band interference filters, 340, 405, 492, 510, 546, 578, 620 nm.
- Halogen lamp.
- Linear range up 2.5 Abs. – resolution of 0.001 Abs.
- Micro- flow-cell, 50 µl total volume
- Temperature controlled at 37 °C ± 0.1°C by Peltier effect
SAMPLING
- High precision Diluter Module
- Sample aspiration from 3 to 99 µl with 1 µl increments
- Reagent dispensing from 5 µl to 500 µl
- Minimum reaction volume 450 µl
REACTION TRAY
114 disposable reaction cells (6 segments of 19 cells each) temperature controlled at 37 °C.
SOFTWARE MAIN FEATURES
Vegasys software runs Under Windows XP and manages, the following functions:
- Unlimited work list
- Historical patient and calibration database
- Quality control graphs visualization and print-out
- Visualization and print-out of calibration graphs
- Methods programming (open system)
- Management of unwanted sequences
- Database for reagents, standards and controls
- Sample pre-dilution
- Dedicated Software for VET testing
- Multipoint calibration curves and relative graphs
- Bidirectional Interface with Host Computer
- Auto Diagnosis program