This is a simultaneous oxygen, nitrogen & hydrogen elemental analyzer with high accuracy and repeatability suiting to cutting-edge technology.

rightMORE INFORMATION

  • The sample is placed in a graphite crucible and the crucible is maintained between the upper and the lower electrode of the impulse furnace. A high current power passes through the crucible to create a high temperature.
     
    The oxide in the sample reacts with the graphite crucible, and is extracted as carbon monoxide (CO), and carried together with the carrier gas. The extracted gas is directly analysed after the dust filter. According to the oxygen concentration, it is determined by CO or CO2 after pass throw oxidizer (Copper Oxide) with a non dispersive Infra-Red analyser (NDIR).
     
    The Hydrogen is determined with a non dispersive Infra Red analyser as H2O after H2 passes through the oxidizer (Copper Oxide).

  • Principle
    Oxygen: Non Dispersive Infrared detector
    Nitrogen: Thermal Conductivity detector(TCD)
    Hydrogen: Non dispersive Infrared detector (as H2O)
    Measurement range*:
    Oxygen: 0-5%, Nitrogen: 0-3%, Hydrogen: 0-0.25%
    *Up to 100% wt is possible by decreasing the sample weight.
    Sample weight: 1.0 ± 0.1g
    Sensitivity(Minimum reading): Oxygen/Nitrogen/Hydrogen: 0.001 ppm.
    Accuracy**(Repeatability)
    Oxygen/Nitrogen
    σn-1≦0.02 ppm or RSD≦0.5% whichever is larger (Reference gas)
    σn-1≦0.3 ppm or RSD≦1.0% whichever is larger (Standard sample)
    Hydrogen
    σn-1≦0.04 ppm or RSD≦2% whichever is larger (Reference gas)
    ** Based on the sample weight 1.0 ± 0.1g and with automation functions described below
    Display
    1) Measurement result: LCD or printout
    2) Alarm message: LCD or printout
    3) Flow sheet: LCD
    Power of furnace
    Inert gas fusion in impulse furnace
    Enable to set the generating power 0 – 8.0 kw
    Sample introduction: Dual sample/flux introduction mechanism
    Automation functions: Auto cleaner, Crucible loader
    Integration conditions:
    Preset time integration
    Integration up to reach comparator level
    Preset time integration or integration up to reach comparator level, whichever is shorter
    Sample ID: Enter up to 20 one-byte characters
    Calibration
    One point or multi point calibration (Reference gas or standard samples)
    Calibration using previous analysis data
    Calibration curve correction function
    Functions
    Display of real time extraction curve
    Analysis interruption
    Self diagnosis and alarm display
    Analysis of extraction curve (Superposition, difference etc analysis)
    Out put (RC-232C or TCP/IP)
    Dimensions: 750mm(H)×785mm(D)×653mm(W)
    Sample introduction is positioned at 650mm from table.
    Weight: 230kg
    On carrying, it is divided in two units (Under 140kg/unit)
    Computer: PC model operating OS Windows® XP(SP2) and beyond
    Power
    Main unit: AC200/220/230/240V ± 10%
    Auto cleaner: AC100V (Step-down transformer)
    Frequency: 50/60 ± 1Hz
    Electric power consumption
    Main unit:12kVA (MAX)
    Auto cleaner: 1.5kVA (MAX)
    Ground resistance: Less than 10Ω
    Installation condition
    Operation temperature: 5-40°C
    Performance temperature: 5-35°C
    Humidity: Maximal relative humidity 80%RH between 5-31℃, linearly decrease down to 50%RH between 31-40℃
    Vibration: Duplex amplitude 20micron and less than 0.098m/S2 accelerations at frequency band
    Required gas
    He carrier gas: Purity greater than 99.995%, Pressure 0.35MPa
    Stainless steel pipe(O.D.3mm) and suitable connecter fitting within 3m of unit
    Dry air or N2 as operating gas: Pressure 0.45MPa
    Nylon pipe (O.D.6mm) and suitable connecter fitting within 5m of unit
    Cooling mechanism: Radiator unit (Separation type)
    Electronic balance (Option)
    Enable connection with electronic balance with 1-0.01mg minimum display
    Optional automatic voltage regulator (AVR)
    Capacity:15kVA
    Weight: 130kg

 