Principle
Oxygen: Non Dispersive Infrared detector
Nitrogen: Thermal Conductivity detector(TCD)
Hydrogen: Non dispersive Infrared detector (as H2O)
Measurement range*:
Oxygen: 0-5%, Nitrogen: 0-3%, Hydrogen: 0-0.25%
*Up to 100% wt is possible by decreasing the sample weight.
Sample weight: 1.0 ± 0.1g
Sensitivity(Minimum reading): Oxygen/Nitrogen/Hydrogen: 0.001 ppm.
Accuracy**(Repeatability)
Oxygen/Nitrogen
σn-1≦0.02 ppm or RSD≦0.5% whichever is larger (Reference gas)
σn-1≦0.3 ppm or RSD≦1.0% whichever is larger (Standard sample)
Hydrogen
σn-1≦0.04 ppm or RSD≦2% whichever is larger (Reference gas)
** Based on the sample weight 1.0 ± 0.1g and with automation functions described below
Display
1) Measurement result: LCD or printout
2) Alarm message: LCD or printout
3) Flow sheet: LCD
Power of furnace
Inert gas fusion in impulse furnace
Enable to set the generating power 0 – 8.0 kw
Sample introduction: Dual sample/flux introduction mechanism
Automation functions: Auto cleaner, Crucible loader
Integration conditions:
Preset time integration
Integration up to reach comparator level
Preset time integration or integration up to reach comparator level, whichever is shorter
Sample ID: Enter up to 20 one-byte characters
Calibration
One point or multi point calibration (Reference gas or standard samples)
Calibration using previous analysis data
Calibration curve correction function
Functions
Display of real time extraction curve
Analysis interruption
Self diagnosis and alarm display
Analysis of extraction curve (Superposition, difference etc analysis)
Out put (RC-232C or TCP/IP)
Dimensions: 750mm(H)×785mm(D)×653mm(W)
Sample introduction is positioned at 650mm from table.
Weight: 230kg
On carrying, it is divided in two units (Under 140kg/unit)
Computer: PC model operating OS Windows® XP(SP2) and beyond
Power
Main unit: AC200/220/230/240V ± 10%
Auto cleaner: AC100V (Step-down transformer)
Frequency: 50/60 ± 1Hz
Electric power consumption
Main unit:12kVA (MAX)
Auto cleaner: 1.5kVA (MAX)
Ground resistance: Less than 10Ω
Installation condition
Operation temperature: 5-40°C
Performance temperature: 5-35°C
Humidity: Maximal relative humidity 80%RH between 5-31℃, linearly decrease down to 50%RH between 31-40℃
Vibration: Duplex amplitude 20micron and less than 0.098m/S2 accelerations at frequency band
Required gas
He carrier gas: Purity greater than 99.995%, Pressure 0.35MPa
Stainless steel pipe(O.D.3mm) and suitable connecter fitting within 3m of unit
Dry air or N2 as operating gas: Pressure 0.45MPa
Nylon pipe (O.D.6mm) and suitable connecter fitting within 5m of unit
Cooling mechanism: Radiator unit (Separation type)
Electronic balance (Option)
Enable connection with electronic balance with 1-0.01mg minimum display
Optional automatic voltage regulator (AVR)
Capacity:15kVA
Weight: 130kg